Specifications Spectrocam 75 RE
With its unique patent pending design, the Spectrocam matches or even exceeds specifications of high end spectrophotometers at a fraction of the price.
As one of the few spectrographic meters on the market the Spectrocam meets ISO standards for color measurement. At the same time the Spectrocam measures many times faster and is more versatile in use than any other handheld spectrophotometers on the market.
Included with the Spectrocam is a ISO standard black scan plate and scan ruler for measuring color bars allowing you to measure 18 targets in less than 10 seconds. It also includes a white reference and monitor accessory designed to calibrate both CRT and LCD monitors in 30 seconds.
Spectrophotometer Module
|
Aperture: |
1.5 x2 mm |
|
|
ISO standard geometry: |
0° (±5°) |
|
|
Spectral range: |
380-750 nm |
|
|
Optical resolution: |
5nm (75 bands) |
|
|
Bandwidth: |
20 nm |
|
|
Wavelength calibration: |
< 2 nm |
|
|
Spectral reporting: |
5nm Interval |
|
|
Analog-to-Digital converter: |
12 bits |
Reflection Illuminator Module
|
Illuminated aperture: |
2 options: 3.5x4mm and 5.5x6mm |
|
|
ISO standard geometry: |
Annular 45° |
|
|
Illumination source: |
Pulsed Xenon gas discharge tube |
|
|
Maximum pulse frequency: |
35 Hz |
|
|
Reference measuring: |
Light output, Voltage, Temperature |
Software Functions
|
Color spaces: |
CIE XYZ, xyY, Luv, Lab, LCH, correlated color temperature, whiteness index, HSB, RGB |
|
|
Color difference values: |
|
|
|
Observers: |
CIE 2° and 10° human observer |
|
|
Illuminants: |
A, C, E, D50, D55, D65, D75 |
|
|
Density functions: |
Status A, T, E, I and DIN 16536-2 |
Dimensions And Accessories
|
Spectrocam: |
130(l) x 70(w) x 50(h) mm |
|
| Available colors: | arctic Blue, Mauve | |
|
Weight excluding accessories: |
180 grams |
|
|
Scan plate: |
335(l) x 240(w) x 6(h) mm |
|
|
Scan plate surface: |
ISO standard black backing material |
|
|
Scan guide: |
250 mm scan length |
|
|
Monitor accessory: |
For use on all screen types (no suction cups) |
|
|
Serial connection plugs: |
Apple Mac modem port; IBM PC RS 232 sub-D9 |
|
|
AC adapter: |
90-260 VAC, 50/60 Hz to 12 VDC, 0.5 A. |
PERFORMANCE
Scan Function
|
Number of readings per patch: |
Averages 4 to 16 full spectral readings |
|
|
Average scan time per patch: |
0.4 sec. For screen frequencies >= 85 lines/inch |
|
|
Minimum screen frequency: |
45 lines/inch (=1.8 lines/mm). Scan time 2 sec |
|
|
Target dimensions: |
Recommended 10x10mm. Minimum7x8 mm |
Spectro Colorimeter
| Reflectance range: |
|
0 to 160% |
| Typical measuring time: |
|
0.4 sec (based upon 12 readings) |
| Short-term repeatability on white: |
|
Typ. 0.1 |
| lnterinstrument accuracy: |
|
Typ 0.4 dE (max. 0.8 dE) |
| White calibration traceable to: |
|
European Community Bureau of Reference |
Spectro densitometer
|
Density range: |
-0.200 to + 3.00 |
|
|
Measuring time: |
Typ. 0.4 sec (based upon 12 readings) |
|
|
Short-term repeatability on white: |
Typ. 0.002 (max. 0.004) |
|
|
Linearity: |
Typ. 0.003 (max. 0.005) at D=0.500 |
Monitor Calibrator*
|
Range: |
0 to 100% |
|
|
Measuring aperture: |
6mm |
|
|
Typical calibration time |
30 seconds |
|
|
Screen synchronization: |
50 to 250 Hz |
|
|
Resolution: |
0.03 % |
|
|
Noise: |
Typ. 0.1 % |
Spectral Camera*
|
Field of view: |
<= 10° |
|
|
Shutter speeds: |
1/2000 to 1 second |
|
|
Data transmission rate: |
Up to 35 spectral images per second |
*not available in the US market in combination with reflective version, due to patent restrictions
Ordering Information Introduction
Last Updated: 23-12-03